SECTION 6<br />

Principle of Operation<br />

The schematic illustration on the following page depicts the<br />

Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer operations. The<br />

primary operations are described below.<br />

Fluid Warming<br />

The Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer utilizes a<br />

solution reservoir housed in a controller unit. Recirculating solution is<br />

warmed and pumped through a heat exchanger (a part of the<br />

Disposable Set). The solution is returned to the reservoir for<br />

continuous recirculation and remains isolated from the patient and<br />

from the irrigating fluid path. The recirculating solution is heated to a<br />

pre-set temperature. The system continuously monitors and controls<br />

the recirculating solution temperature. The Level 1 ® NORMOFLO ®<br />

H-<strong>1100</strong> Irrigation Fluid Warmer is designed to stop fluid warming and<br />

provide audible and visual alarms in the event of an over-temperature<br />

condition.<br />

S E C T I O N 6 • P r i n c i p l e o f O p e r a t i o n<br />

12 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual