Slow Flow Rate Troubleshooting<br />
Problem Check the following<br />
Fluid bag not fully<br />
spiked or bag port<br />
twisted<br />
Clamps partly<br />
engaged<br />
Clamps left in the<br />
clamped position for<br />
long periods<br />
S E C T I O N 1 0 • T r o u b l e s h o o t i n g<br />
Make sure the membrane of the fluid bag port is fully<br />
pierced by the bag spike and that the neck of the bag<br />
port is not twisted.<br />
Verify all clamps are fully open.<br />
Leaving clamps fully clamped for long periods will cause<br />
the tubing to become deformed. Do not leave clamps<br />
closed for extended periods of time.<br />
Kinked tubing Verify that no tubing kinks are present.<br />
Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual 31