Slow Flow Rate Troubleshooting<br />

Problem Check the following<br />

Fluid bag not fully<br />

spiked or bag port<br />

twisted<br />

Clamps partly<br />

engaged<br />

Clamps left in the<br />

clamped position for<br />

long periods<br />

S E C T I O N 1 0 • T r o u b l e s h o o t i n g<br />

Make sure the membrane of the fluid bag port is fully<br />

pierced by the bag spike and that the neck of the bag<br />

port is not twisted.<br />

Verify all clamps are fully open.<br />

Leaving clamps fully clamped for long periods will cause<br />

the tubing to become deformed. Do not leave clamps<br />

closed for extended periods of time.<br />

Kinked tubing Verify that no tubing kinks are present.<br />

Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual 31