Corrective Action<br />
New and old plates must exhibit bacteria growth. If new plates do not exhibit bacteria growth,<br />
but the old plates do, then the batch <strong>of</strong> new plates is assumed to be unable to support bacteria<br />
and discarded. If both the new and old plates do not exhibit bacteria growth then it is assumed<br />
that the plates were not inoculated properly and the inoculation procedure is repeated.<br />
Notes regarding dilutions for QA samples<br />
Volume <strong>of</strong><br />
sample<br />
filtered<br />
(mL)<br />
Dilution tube preparation Volume<br />
placed on<br />
filter (mL)<br />
Calculated Result<br />
(# <strong>of</strong> colonies divided<br />
by calculated result =<br />
colonies/100 mL)<br />
0.1 1 mL E. coli inoculated TSB sample<br />
into 9 mL <strong>of</strong> PBS<br />
1 1,000 or 10 -3<br />
0.01 1 mL 10 3 sample into 9 mL <strong>of</strong> PBS 1 10,000 or 10 -4<br />
0.001 1 mL 10 4 sample into 9 mL <strong>of</strong> PBS 1 100,000 or 10 -5<br />
0.0001 1 mL 10 5 sample into 9 mL <strong>of</strong> PBS 1 1,000,000 or 10 -6<br />
0.00001 1 mL 10 6 sample into 9 mL <strong>of</strong> PBS 1 10,000,000 or 10 -7<br />
0.000001 1 mL 10 7 sample into 9 mL <strong>of</strong> PBS 1 100,000,000 or 10 -8<br />
5.3 Analysis Method<br />
5.3.1 Preparation – 1 Week Before Scheduled Sampling<br />
1. Check to be sure the UV box works. If the lights flash and turn on when the black button<br />
at the left end <strong>of</strong> the box is depressed – it works. Be sure not to stare at the light as it<br />
can burn your retinas! If it doesn’t work contact Elizabeth Herron to have the box<br />
repaired.<br />
2. Check to be sure there is an adequate supply <strong>of</strong> mTEC media plates that are less than 1<br />
month old. If there are not enough plates, more must be prepared. Preparation<br />
instructions are found in Section 5.3.1.1.<br />
Materials<br />
5.3.1.1 Preparation <strong>of</strong> mTEC Media Plates<br />
Envirocide or Conflict Magnetic stirring bar<br />
Sterile Petri dishes Magnetic stirring and heating plate<br />
Autoclave Metal tray<br />
2 - 2 L autoclave safe flasks Aluminum foil<br />
Insulated autoclave gloves Sterile 25 mL pipette<br />
Dehydrated mTEC media Empty foil lined cardboard box<br />
Deionized (DI) water Electronic pipette-aid<br />
9 <strong>of</strong> 23<br />
Ambient Waters Microbiological Procedure SOP 007<br />
S:\WW\awwword\LABPROC\all QAPPs\LABQAPPs\QAPP Rev5 -0609\SOPs\SOP 007AmbientMicrobes.doc