URI Watershed Watch Laboratory<br />

Buffered Dilution/rinse Water KH 2PO 4 reagent pH Record<br />

Tech's<br />

Initials<br />

Meter<br />

Letter<br />

(A or B)<br />

Calibration 2<br />

Intitial pH<br />

Volume 1N<br />

NaOH added<br />

(mL)<br />

Final pH Comments / Notes<br />

This form is intended to be used to maintain a record <strong>of</strong> pH and adjustment for the phosphate reagent for the fecal coliform dilution/rinse<br />

water. When new reagent is made up, the initial pH, amount <strong>of</strong> NaOH added, and final pH should be recorded on this sheet. Once this<br />

sheet is full it should be replaced with a new blank sheet. The full sheet should be returned to the URIWW Micro Laboratory Binder in<br />

Room 019; blank sheets are also located in this binder. Check the Laboratory QAPP, SOP 010 - Alkalinity and pH procedures, for<br />

details regarding the pH and alkalinity procedures.<br />

Calibration and Maintenance Information:<br />

1<br />

Each time the pH probe is used the filling solution level in the probe should be checked. The level should be within 1/4 inch <strong>of</strong> the top <strong>of</strong><br />

the electode. If it isn't, add enough filling solution to bring it up to that point.<br />

2 The meter should have a slope (Electrode Efficiency) <strong>of</strong> at least 96% after calibration. If the meter doesn't initially calibrate correctly,<br />

change the probe filling solution and replace the pH buffers, then recalibrate the probe. If this doesn't solve the problem replace the<br />

probe. See Linda or Elizabeth with questions.<br />

Ambient Waters Microbiological Procedure SOP 007<br />

S:\WW\awwword\LABPROC\all QAPPs\LABQAPPs\QAPP Rev5 -0609\SOPs\SOP 007AmbientMicrobes.doc