2.1.2 Annual Metals Testing<br />

Annually, water obtained from the Aries Vaponics units in Rooms 002 and Room 019 will be<br />

tested for lead, cadmium, chromium, copper, nickel and zinc. These contaminations will be at a<br />

level <strong>of</strong> less than 0.05 mg/L per contaminant and less than 0.1 mg/L when added together.<br />

Corrective Action<br />

If the annual metals testing reveals metal value(s) at a level greater than acceptable then the<br />

water will not be used for microbiological purposes until another sample is taken and returns an<br />

acceptable value.<br />

2.1.3 Monthly heterotrophic plate count<br />

Monthly a sample <strong>of</strong> water from Room 002 and 019 will be analyzed for heterotrophic plate<br />

count. The test results will be less than 500 CFU/ml.<br />

Corrective Action<br />

If the monthly heterotrophic plate count(s) return a value greater than 500 CFU/ml then the<br />

affected water source(s) will not be used for microbiological purposes until another sample is<br />

taken and returns an acceptable value.<br />

Laboratory Water SOP 002<br />

S:\WW\awwword\LABPROC\all QAPPs\LABQAPPs\QAPP Rev5 -0609\SOPs\SOP 002 Lab Water-MOD.doc