OSMOMAT 3000<br />
Operation<br />
Fig. 37: Pipette sample<br />
Fig. 38: Position measuring vessel<br />
Fig. 39: Measurement menu<br />
50 Version 1.02 04/24/2013<br />
6. Pipette a sample volume of 50µl into an unused and clean measuring<br />
vessel (Fig. 37).<br />
NOTE!<br />
Risk of incorrect measurement resulting from the<br />
presence of air bubbles in the sample!<br />
The sample must be pipetted without air bubbles.<br />
7. Position measuring vessel on thermistor probe with cover facing<br />
front (Fig. 38/1).<br />
8. Option D only:<br />
The measuring vessel is securely attached to the<br />
thermistor probe when it clicks into place.<br />
Tap "Printer On" on the measurement menu (Fig. 39/1) to output<br />
the measurement results to a printer. Make sure that the printer is<br />
ready.<br />
9. To start the measurement, tap "Single Sample" (Fig. 39/2).<br />
If necessary, enter the sample ID using the virtual keyboard.<br />
The sample ID can be pre-defined in the user preferences.<br />
In this case, the system assigns the sample<br />
ID automatically ( page 63)