OSMOMAT 3000<br />

Operation<br />

Fig. 37: Pipette sample<br />

Fig. 38: Position measuring vessel<br />

Fig. 39: Measurement menu<br />

50 Version 1.02 04/24/2013<br />

6. Pipette a sample volume of 50µl into an unused and clean measuring<br />

vessel (Fig. 37).<br />

NOTE!<br />

Risk of incorrect measurement resulting from the<br />

presence of air bubbles in the sample!<br />

The sample must be pipetted without air bubbles.<br />

7. Position measuring vessel on thermistor probe with cover facing<br />

front (Fig. 38/1).<br />

8. Option D only:<br />

The measuring vessel is securely attached to the<br />

thermistor probe when it clicks into place.<br />

Tap "Printer On" on the measurement menu (Fig. 39/1) to output<br />

the measurement results to a printer. Make sure that the printer is<br />

ready.<br />

9. To start the measurement, tap "Single Sample" (Fig. 39/2).<br />

If necessary, enter the sample ID using the virtual keyboard.<br />

The sample ID can be pre-defined in the user preferences.<br />

In this case, the system assigns the sample<br />

ID automatically ( page 63)