Important Information – Please Read Before Use<br />
• Do not touch the electrical contacts inside the electrical<br />
connector. CCD damage may result.<br />
• Do not apply shock to the distal end of the insertion tube,<br />
particularly the objective lens surface at the distal end. Visual<br />
abnormalities may result.<br />
• Do not twist or bend the bending section with your hands.<br />
Equipment damage may result.<br />
• Do not squeeze the bending section forcefully. The covering<br />
of the bending section may stretch or break and cause water<br />
leaks.<br />
• Turn the video system center OFF before connecting or<br />
disconnecting the videoscope cable from the electrical<br />
connector on the endoscope. Turn the video system center<br />
ON or OFF only when the videoscope cable is connected to<br />
both the video system center and electrical connector on the<br />
endoscope. Failure to do so can result in equipment damage,<br />
including destruction of the CCD.<br />
• The endoscope’s remote switches cannot be removed from<br />
the control section. Pressing, pulling or twisting them with<br />
excessive force can break the switches and/or may cause<br />
water leaks.<br />
• If remote switch 1 does not return to the OFF position after<br />
being pressed strongly from the side, gently pull the switch<br />
upwards to return it to the OFF position.<br />
• Do not hit or bend the electrical contacts on the endoscope<br />
connector. The connection to the light source may be<br />
impaired and faulty contact can result.<br />
• Do not attempt to bend the endoscope’s insertion tube with<br />
excessive force. Otherwise, the insertion tube may be<br />
damaged.<br />
• Electromagnetic interference may occur on this instrument<br />
near equipment marked with the following symbol or other<br />
portable and mobile RF (Radio Frequency) communications<br />
equipment such as cellular phones. If electromagnetic<br />
interference occurs, mitigation measures may be necessary,<br />
such as reorienting or relocating this instrument, or shielding<br />
the location.<br />
EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />
7