Chapter 3 Preparation and Inspection<br />
1. Confirm that the slits and holes on the biopsy valves have no splits, cracks,<br />
deformation, discoloration or other damage (see Figure 3.7).<br />
Normal<br />
Outside of Cap<br />
Abnormal<br />
Splits, Cracks<br />
Slit<br />
Inside of Cap<br />
Discoloration<br />
Main Body<br />
Hole<br />
Splits, Cracks<br />
Figure 3.7<br />
2. Firmly install two cap parts on the main body as shown in Figure 3.8.<br />
Be sure to push the suction switchover lever all the way into<br />
the biopsy valve. Otherwise, when turning the suction<br />
switchover, the suction instrument channels cannot be<br />
changed correctly.<br />
EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />
31