Chapter 3 Preparation and Inspection<br />
1. Confirm that the mouthpiece is free from cracks, deformation or<br />
discoloration (see Figure 3.9).<br />
Opening<br />
Main body<br />
Outer flange<br />
Figure 3.9<br />
2. Using your fingers, check for excessive scratching or other irregularities on<br />
all surface of the mouthpiece (see Figure 3.9).<br />
3.4 Attaching accessories to the endoscope<br />
The air/water valve and the suction valve do not require<br />
lubrication. Lubricants can cause swelling of the valves’<br />
seals, which will impair valve function.<br />
Attaching the suction valve<br />
1. Align the two metal ridges on the underside of the suction valve with the two<br />
holes in the suction cylinder.<br />
2. Attach the suction valve to the suction cylinder of the endoscope (see<br />
Figures 3.10 and 3.11). Confirm that the valve fits properly without any<br />
bulging of the skirt and that the valve does not rotate.<br />
EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />
33