Chapter 4 Operation<br />
Chapter 4<br />
Operation<br />
The operator of this instrument must be a physician or medical personnel under<br />
the supervision of a physician and must have received sufficient training in<br />
clinical endoscopic technique. This manual, therefore, does not explain or<br />
discuss clinical endoscopic procedures. It only describes basic operation and<br />
precautions related to the operation of this instrument.<br />
• Wear personal protective equipment to guard against<br />
dangerous chemicals and potentially infectious material<br />
during operation. During operation, wear appropriate<br />
personal protective equipment, such as eye wear, face mask,<br />
moisture-resistant clothing, chemical-resistant gloves that fit<br />
properly and are long enough so that your skin is not<br />
exposed.<br />
• The temperature of the distal end of the endoscope may<br />
exceed 41°C (106°F) and reach 50°C (122°F) due to intense<br />
endoscopic illumination. Surface temperatures over 41°C<br />
(106°F) may cause mucosal burns. Always use the minimum<br />
level of illumination, minimum time and suitable distance<br />
necessary for adequate viewing. Whenever possible, avoid<br />
close stationary viewing and do not leave the distal end of the<br />
endoscope close to the mucous membrane for a long time.<br />
• Whenever possible, do not leave the endoscope illuminated<br />
before and/or after an examination. Continued illumination<br />
will cause the distal end of the endoscope to become hot and<br />
could cause operator and/or patient burns.<br />
• Turn the video system center ON to operate the light source’s<br />
automatic brightness function. When the video system center<br />
is OFF, it cannot operate the light source’s automatic<br />
brightness function, and the light intensity is set to the<br />
maximum level. In this case, the distal end of the endoscope<br />
can become hot and could cause operator and/or patient<br />
burns (when using a light source expect CLV-260).<br />
EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />
45