Chapter 4 Operation<br />

Chapter 4<br />

Operation<br />

The operator of this instrument must be a physician or medical personnel under<br />

the supervision of a physician and must have received sufficient training in<br />

clinical endoscopic technique. This manual, therefore, does not explain or<br />

discuss clinical endoscopic procedures. It only describes basic operation and<br />

precautions related to the operation of this instrument.<br />

• Wear personal protective equipment to guard against<br />

dangerous chemicals and potentially infectious material<br />

during operation. During operation, wear appropriate<br />

personal protective equipment, such as eye wear, face mask,<br />

moisture-resistant clothing, chemical-resistant gloves that fit<br />

properly and are long enough so that your skin is not<br />

exposed.<br />

• The temperature of the distal end of the endoscope may<br />

exceed 41°C (106°F) and reach 50°C (122°F) due to intense<br />

endoscopic illumination. Surface temperatures over 41°C<br />

(106°F) may cause mucosal burns. Always use the minimum<br />

level of illumination, minimum time and suitable distance<br />

necessary for adequate viewing. Whenever possible, avoid<br />

close stationary viewing and do not leave the distal end of the<br />

endoscope close to the mucous membrane for a long time.<br />

• Whenever possible, do not leave the endoscope illuminated<br />

before and/or after an examination. Continued illumination<br />

will cause the distal end of the endoscope to become hot and<br />

could cause operator and/or patient burns.<br />

• Turn the video system center ON to operate the light source’s<br />

automatic brightness function. When the video system center<br />

is OFF, it cannot operate the light source’s automatic<br />

brightness function, and the light intensity is set to the<br />

maximum level. In this case, the distal end of the endoscope<br />

can become hot and could cause operator and/or patient<br />

burns (when using a light source expect CLV-260).<br />

EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />

45