Chapter 3 Preparation and Inspection<br />
Inspection of the auxiliary water feeding function<br />
Use sterile water only. Using non-sterile water may cause<br />
patient cross-contamination and infection.<br />
1. Attach a syringe containing sterile water or the water tube from a water<br />
pump to the luer port of the auxiliary water tube (see Figure 3.19). Feed<br />
water and confirm that water is emitted from the auxiliary water channel at<br />
the distal end of the insertion tube.<br />
2. Make sure that no water leaks at the connection between the connecting<br />
end of the auxiliary water tube and the auxiliary water inlet.<br />
3. Make sure that no water leaks at the connection between the luer port of the<br />
auxiliary water tube and the syringe or the water tube.<br />
4. Disconnect the water tube from the water pump or the syringe from the luer<br />
port of the auxiliary water tube. Make sure that no water leaks at the luer<br />
port of the auxiliary water tube and/or the distal end of the insertion tube.<br />
• If the auxiliary water channel is used for feeding water, never<br />
disconnect the auxiliary water tube during examination, leave<br />
it attached until the endoscope is precleaned. If the auxiliary<br />
water tube is detached before precleaning, water remaining<br />
in the auxiliary water channel may be spilled on the<br />
surrounding equipment. This could cause damage and/or<br />
malfunction of the equipment.<br />
• Note that the luer port on the MAJ-855 includes a one-way<br />
valve to prevent backflow – do not use the MAJ-855 without<br />
this connector in place, otherwise backflow of contaminated<br />
material may occur and equipment damage or patient injury<br />
may result.<br />
EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />
43