Chapter 3 Preparation and Inspection<br />

3.3 Preparation and inspection of accessories<br />

Clean and disinfect or sterilize the air/water valve, suction valve, biopsy valve<br />

and auxiliary water tube as described in the “OLYMPUS GIF TYPE 2TQ260M<br />

REPROCESSING MANUAL”.<br />

Inspection of the air/water and suction valves<br />

Confirm that the top hole of the air/water valve is not blocked.<br />

If the hole is blocked, air is fed continuously and patient pain,<br />

bleeding and/or perforation can result.<br />

1. Confirm that the holes of the valves are not blocked (see Figures 3.5 and<br />

3.6).<br />

2. Confirm that the valves are not deformed or cracked (see Figures 3.5 and<br />

3.6).<br />

3. Check for excessive scratching or tears in the air/water valve’s seals (see<br />

Figure 3.5).<br />

Spring<br />

Hole<br />

Seals<br />

Skirt<br />

Hole<br />

Air/water valve (MH-438)<br />

Figure 3.5<br />

Spring<br />

Skirt<br />

Hole<br />

Suction valve (MH-443)<br />

Figure 3.6<br />

EVIS LUCERA GIF TYPE 2TQ260M OPERATION MANUAL<br />

29