eing c<strong>on</strong>nected to a patient.<br />

2. C<strong>on</strong>figure SpO 2 alarms: Pulse oximetry alarms can be manually set.<br />

a. Touch <strong>the</strong> Settings tab.<br />

b. Touch to select M<strong>on</strong>itor.<br />

c. Touch <strong>the</strong> Alarms tab.<br />

d. Touch SpO 2 .<br />

e. To turn alarm <strong>on</strong> or off, use <strong>the</strong> up and down arrow keys or <strong>the</strong> numeric<br />

keypad to adjust upper and lower SpO 2 alarm limits.<br />

f. Touch Home to save your settings, return to <strong>the</strong> Home tab, and take an SpO 2<br />

measurement.<br />

WARNING: Inspect <strong>the</strong> SpO 2 cable. Replace it if it shows any signs of wear,<br />

breakage or fraying. Failure to do so could harm <strong>the</strong> patient, harm <strong>the</strong> m<strong>on</strong>itor, or<br />

produce inaccurate readings.<br />

NOTE:<br />

• During an SpO 2 measurement, <strong>the</strong> displayed pulse rate is derived from <strong>the</strong><br />

SpO 2 sensor. If SpO 2 is not available, <strong>the</strong> pulse rate is derived from NIBP.<br />

• If SpO 2 is being measured c<strong>on</strong>tinuously <strong>on</strong> a patient for an extended period,<br />

change <strong>the</strong> sensor locati<strong>on</strong> at least every three hours or as indicated by <strong>the</strong><br />

sensor manufacturer's directi<strong>on</strong>s for use.<br />

3. SatSec<strong>on</strong>ds alarm management: The SatSec<strong>on</strong>ds feature is an SpO 2 alarm<br />

management system available <strong>on</strong>ly with Nellcor-equipped m<strong>on</strong>itors.<br />

a. The SatSec<strong>on</strong>ds feature has a built-in safety protocol that sounds an alarm<br />

whenever three SpO 2 violati<strong>on</strong>s of any amount or durati<strong>on</strong> occur within a <strong>on</strong>eminute<br />

period.<br />

b. The SatSec<strong>on</strong>ds feature delays alarms for a set amount of time when it is<br />

enabled. If a c<strong>on</strong>diti<strong>on</strong> or alarm resolves <strong>on</strong> its own within <strong>the</strong> preset time, <strong>the</strong><br />

alarm will automatically reset.<br />

c. The feature can be set to 0, 10, 25, 50, or 100 SatSec<strong>on</strong>ds. The SatSec<strong>on</strong>ds<br />

feature is <strong>the</strong> product of <strong>the</strong> time and magnitude that a patient exceeds SpO 2<br />

alarm limits. For example, three points below <strong>the</strong> alarm limit for 10 sec<strong>on</strong>ds<br />

equals 30 SatSec<strong>on</strong>ds. An alarm is triggered <strong>on</strong>ly when a desaturati<strong>on</strong> event<br />

reaches <strong>the</strong> SatSec<strong>on</strong>ds limit.<br />

d. When <strong>the</strong> feature is not enabled, <strong>the</strong> SatSec<strong>on</strong>ds graphic will not appear and<br />

SpO 2 alarm c<strong>on</strong>diti<strong>on</strong>s will not be delayed.<br />

e. To set SatSec<strong>on</strong>ds limits:<br />

i. Touch <strong>the</strong> Alarms limit c<strong>on</strong>trol of <strong>the</strong> SpO 2 frame.<br />

ii. Touch <strong>the</strong> Alarms tab.<br />

iii. Touch to select a SatSec<strong>on</strong>ds setting. Values are 10, 25, 50, or 100.<br />

iv. Touch Home to save your settings and return to <strong>the</strong> Home tab.<br />

4. Perfusi<strong>on</strong> index: Perfusi<strong>on</strong> index is an SpO 2 feature available <strong>on</strong>ly with Masimoequipped<br />

m<strong>on</strong>itors.<br />

a. Perfusi<strong>on</strong> Index (PI) is a relative reading of pulse strength at <strong>the</strong> m<strong>on</strong>itoring<br />

site and is a numerical value that indicates <strong>the</strong> strength of <strong>the</strong> IR (infrared)<br />

C<strong>on</strong>nex Vital Signs M<strong>on</strong>itor 6000 Trainer Outline<br />

MC 7517 08/2010<br />

9