26.08.2014 Views

Section 6: Selected Applications & Techniques

UT testing self study notes

UT testing self study notes

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2.0 Methods used for H.Temperature Scanning<br />

2.1 Transducers- H.Temperature Delay Line Material<br />

Panametrics-NDT high temperature transducers fall into two categories,<br />

■<br />

■<br />

dual element transducers and<br />

delay line transducers.<br />

In both cases, the delay line material (which is internal in the case of duals)<br />

serves as thermal insulation between the active transducer element and the<br />

hot test surface.<br />

For design reasons, there are no high temperature contact or immersion<br />

transducers in the standard product line. High temperature duals and delay<br />

line transducers are available for both thickness gaging and flaw detection<br />

applications. As with all ultrasonic tests, the best transducer for a given<br />

application will be determined by specific test requirements, including the<br />

material, the thickness range, the temperature, and in the case of flaw<br />

detection, the type and size of the relevant flaws.