8.1 External Level Detector<br />

unistat control V3.8.1<br />

The level detector is installed in the open bath, to be thermoregulated. As soon as the level<br />

decreases significantly a dangerous condition exists (a leak?). The level detector connected to<br />

the unistat control passes this on and the unit will be switched off.<br />

For exact description see Appendix D.<br />

8.2 Flow Detector<br />

The flow detector monitors the circulation. If the circulation pressure falls during<br />

thermoregulation<br />

it will release an emergency shut down. (See message 41/X in Appendix B and program 55).<br />

8.3 Fault switches<br />

Depending on the UNISTAT model various sensors will be analysed. Messages inform about<br />

status (see program 10).<br />

8.4 Current Sensor / Power On Self-Test<br />

The current sensor is an important device especially for testing the functions of contactors,<br />

relays and heaters.<br />

By energising such components their function can be evaluated in comparison with the<br />

measured power input.<br />

8.5 Over Heating<br />

Every UNISTAT model is equipped with over heat protection, which operates independently<br />

form the unistat control. Only requires one sensor register excess temperature and the<br />

heating will be switched off immediately. A more comprehensive description is contained in<br />

the operating manual for the particular thermostat.<br />

Chillers without heaters do not have over heat protection.<br />

! The over heat protection should be triggered regularly, by setting the over heat<br />

protection set point temperature just above the present set point and then heat a little.<br />

Only a fully operational over heat protector will help to avoid damage!<br />

8.6 Condensing Temperature Control<br />

During thermoregulation the condensing temperature of the unit is constantly monitored.<br />

When the maximum value is reached the cooling power of the unit will be reduced, reduces<br />

the condenser heat rejection. This condition will be displayed as a message after a few<br />

seconds. The warning will remain, until it is acknowledged (see program 11, „acknowledge<br />

warning„).<br />

Kondens.Temp.!<br />

or<br />

Condens.Temp.!<br />

11<br />

or<br />

T°C condenseur