unistat control V3.8.1<br />

Appendix A - Program Directory<br />

Program 85: Baud Rate for serial data transmission (RS232/485)<br />

The transmission speed with which your unit is communicating with the connected PC, can<br />

be chosen in one of four rates. Enter , the current baud rate appears:<br />

Host Bd-Rate<br />

Akt1.2kBd<br />

With you leave the program without change, with you go on in the selection:<br />

9.6 4.8<br />

2.4 1.2<br />

Once the selection is completed the transmission speed starts straight away. Always<br />

remember that the connected units (in a RS485-bus system) are set at the same transmission<br />

speed and the master (so the PC, which has a control software installed) has to be set at the<br />

same speed.<br />

Behaviour after MAINS OFF:<br />

This setting is not lost after mains off.<br />

Program 86: Selection RS232 ↔ RS485<br />

Two digital serial interface standards are possible with the control unit software, both are<br />

done via the same plug connector (Plug Stv 12). It is possible to work with only one at any<br />

one time. That is why program 86 offers the selection possibilities. This setting remains after<br />

mains off.<br />

Program 90: Language Selection<br />

Program 90 offers the choice between German, English and French. A short reminder of this<br />

program is displayed after mains switch on.<br />

Program 93: Production number/serial number<br />

(control panels from software version 3)<br />

This is the serial number of the control panel.<br />

Program 94: Control Panel Identification<br />

(control panels from software version 3)<br />

A marking, which indicates which printed circuit board is in the control panel.<br />

Program 98: Version number<br />

(Control panels from software version 2.4)<br />

The version number of the software for the control panel appears in the first line. The<br />

information about the controller software version is in the second line.<br />

Program 106: Maximum Temperature Limit (for external Pt sensor control)<br />

If the value set using this program is exceeded for more than 3 seconds an alarm is triggered.<br />

A device message 73/x will be recorded (see Program 10). Note: The default setting for this<br />

value is 320°C, a value that the external Pt100 sensor will never experience, whether it is<br />

connected or not.<br />