unistat control V3.8.1<br />
Nine tempering programs with 30 program segments each are available.<br />
Each tempering program fundamentally consists of the program descriptor and up to<br />
30 program segments. The following settings prevailing at the start of processing a<br />
thermoregulation program will be adopted:<br />
Tempering method (internal or external),<br />
Minimum and maximum set points,<br />
Inclusion of external control signals,<br />
Use of potential free contact,<br />
Use of automatic response to main power failure.<br />
9.2 Program Descriptor<br />
When a tempering process is planned there is a particular path it has to follow the program is<br />
then set to follow this path exactly. The temperature ramps and temperature plateaux are<br />
only part of an automated process. Additionally the following data is required for the program<br />
descriptor at the start of the thermoregulation program. (See program 21).<br />
Start time for the thermoregulation program:<br />
The program start can take place immediately, or whenever you want. The setting for the<br />
start temperature for the first segment takes place in the program descriptor.<br />
A chosen ramp will start from the actual value to the starting temperature. If full power is<br />
required, a high value, for example 20 K/min, has to be set.<br />
Time or temperature stability for the tempering program:<br />
The rule for every program segment is:<br />
If the temperature at the end of the segment is to be checked the ‘temperature stability’<br />
function should be selected. The next program segment will get started, when the actual<br />
value reaches the end temperature of the program segment with fixed accuracy (+/- 1K).<br />
If on completion of the segment time the actual temperature is not to be checked, the time<br />
stable option should be selected.<br />
If you work the tempering program with length ‘0’, details of the starting temperature,<br />
starting ramp, or if it the temperature stable option will be requested. The programmer can<br />
repeat the program segments.<br />
9.3 Program Segments<br />
Each program segment contains two values: The segment end temperature and the duration<br />
of the segment. The starting temperature of the segment results automatically form the end<br />
temperature of the previous segment. The duration of the segment is reduced constantly<br />
during the program run. The next set point is therefore determined from the difference of the<br />
end temperature and the actual value. At the end of a segment it will be checked, if<br />
‘temperature stability’ (See 9.2) has been requested the end temperature is to be reached<br />
with +/- 1K accuracy. If maintenance steps are necessary, the program run time of the<br />
program is extended. The waiting time in the actual segment will be displayed as tw in the<br />
LCD display.<br />
15