unistat control V3.8.1<br />
Appendix A - Program Directory<br />
Program 5: Mains Failure Automatic Function<br />
After mains failure the unistat or Unichiller goes on working with the same parameters as<br />
before. Should there be a main power failure during tempering, the usual self diagnosis tests<br />
are run on start up after restoration of the mains supply. The set tempering is restarted if no<br />
error is present. If this happens during the run of a tempering program (see program 20), it<br />
will continue.<br />
Program 6: POKO max<br />
This is only relevant for set option ‘actual value check’ in program 8. Setting of the upper<br />
temperature level for the potential free contact (POKO) in 0.1K steps relative to the set point:<br />
An entry of 1.5 in this program means, that the POKO stays activated until the actual value is<br />
1.5K above the set point. Factory setting: 1.0K.<br />
Program 7: POKO min<br />
This is only relevant for set option ‘actual value check’ in program 8. Setting of the lower<br />
temperature level for the potential free contact (POKO) in 0.1K steps relative to the set point:<br />
An entry of 1.5 in this program means, that the POKO stays activated until the actual value is<br />
1.5K below the set point. Default setting: 1.0K<br />
Program 8: POKO Programming<br />
There is a range of programming options for the POKO. Pressing the key allows the<br />
operator to browse through the different possibilities. You will find the technical data of the<br />
POKO in appendix D. Important: The OK – status always means the contact is normally<br />
closed.<br />
(0) POKO Function ‘OFF’<br />
The POKO displays the OK – status when the unit is ready to operate. This condition<br />
is after the internal controller check, ca 30sec after the switch on. The OK – status<br />
will be ended by switching off the mains or a fault.<br />
(1) POKO Function ‘Actual Value’<br />
Using the program 6 and 7 you can set the upper and lower limit of the temperature<br />
range around the set point.<br />
The potential free contact displays the status, that the set point and the actual value<br />
differ relative to a pre set temperature range. If the range is exceeded the potential<br />
free contact is switched (away from OK – status), nor will further reactions of the<br />
unistat or Unichiller will take place. If the actual value is in the range, the contact will<br />
be re set in OK – status.<br />
Example: The ‘actual value check’ has been activated. (External control mode).<br />
Using Program 7 ‘POKO max’ a limit of 5K is set and with program 6 ‘POKO min’ 7K<br />
given. The set point is at the moment at 27,5°C. The contact will switch the OK<br />
status off as soon as the external sensor measures more than 32.5°C or less than<br />
20.5°C. It depends on the tempering method set in program 3 whether the external or<br />
the internal sensor is used.<br />