unistat control V3.8.1<br />

Appendix A - Program Directory<br />

Program 5: Mains Failure Automatic Function<br />

After mains failure the unistat or Unichiller goes on working with the same parameters as<br />

before. Should there be a main power failure during tempering, the usual self diagnosis tests<br />

are run on start up after restoration of the mains supply. The set tempering is restarted if no<br />

error is present. If this happens during the run of a tempering program (see program 20), it<br />

will continue.<br />

Program 6: POKO max<br />

This is only relevant for set option ‘actual value check’ in program 8. Setting of the upper<br />

temperature level for the potential free contact (POKO) in 0.1K steps relative to the set point:<br />

An entry of 1.5 in this program means, that the POKO stays activated until the actual value is<br />

1.5K above the set point. Factory setting: 1.0K.<br />

Program 7: POKO min<br />

This is only relevant for set option ‘actual value check’ in program 8. Setting of the lower<br />

temperature level for the potential free contact (POKO) in 0.1K steps relative to the set point:<br />

An entry of 1.5 in this program means, that the POKO stays activated until the actual value is<br />

1.5K below the set point. Default setting: 1.0K<br />

Program 8: POKO Programming<br />

There is a range of programming options for the POKO. Pressing the key allows the<br />

operator to browse through the different possibilities. You will find the technical data of the<br />

POKO in appendix D. Important: The OK – status always means the contact is normally<br />

closed.<br />

(0) POKO Function ‘OFF’<br />

The POKO displays the OK – status when the unit is ready to operate. This condition<br />

is after the internal controller check, ca 30sec after the switch on. The OK – status<br />

will be ended by switching off the mains or a fault.<br />

(1) POKO Function ‘Actual Value’<br />

Using the program 6 and 7 you can set the upper and lower limit of the temperature<br />

range around the set point.<br />

The potential free contact displays the status, that the set point and the actual value<br />

differ relative to a pre set temperature range. If the range is exceeded the potential<br />

free contact is switched (away from OK – status), nor will further reactions of the<br />

unistat or Unichiller will take place. If the actual value is in the range, the contact will<br />

be re set in OK – status.<br />

Example: The ‘actual value check’ has been activated. (External control mode).<br />

Using Program 7 ‘POKO max’ a limit of 5K is set and with program 6 ‘POKO min’ 7K<br />

given. The set point is at the moment at 27,5°C. The contact will switch the OK<br />

status off as soon as the external sensor measures more than 32.5°C or less than<br />

20.5°C. It depends on the tempering method set in program 3 whether the external or<br />

the internal sensor is used.<br />

21