unistat control V3.8.1<br />

Appendix A - Program Directory<br />

‘rrrrr’ gradient of the ramp in k/min or k/h<br />

The gradient of the ramp is made up of the quotient of the temperature difference and the<br />

segment duration and the unit K/h or K/min, depending on given ‘E’.<br />

To process the segment parameters there is a choice between , and<br />

, with and the parameters are set directly, using the<br />

segment parameters are set using algebraic calculations.<br />

requests the entry of the end temperature of the segment. This will be compared<br />

to the unit parameters for minimum and maximum set point. (See program 1 and<br />

2) The value will be rejected if the limit is exceeded. The resolution is 0,1K.<br />

determines the segment duration in a pre determined unit, minute or hour. The<br />

time-set range is per segment 0.1to 3240 min (54 h).<br />

sets the gradient of the ramp in the segment. After setting this question appears:<br />

Fixieren:<br />

End-temp? Time?<br />

After pressing , the time will be set. The new end temperature will be determined by<br />

the ramp that is to be set. The same goes for the fixing of the end temperature when using<br />

. Then the display of the step with changed values for tttt, eeeeee and rrrr will<br />

appear. If unrealistic values for the ramp are entered, the ‘?????’display will appear for the<br />

gradient of the ramp. The result of the entry is 0.1 K/min or 0.1 K/h. Please note that a small<br />

difference between displayed and entered values are possible due to rounding up or down<br />

errors. The entered ramp value is only accepted, if the ranges for the unit temperature range<br />

and the segment duration are not exceeded.<br />

Scrolling Through The Program Segments:<br />

Pressing a segment is switched forward and with backwards.<br />

The programmer must be in the command setting mode to do this, where the keys ‘Ramp’,<br />

’Temp’ and ‘Time’ are to be used.<br />

Naturally the segment number must be within the limits for the smallest and largest segment<br />

number.<br />

If the message ‘sgmt. Activated’, that means that the program length has been extended by<br />

one segment. In this segment the values are shown, which was previously set.<br />

With the previous segment can be called up.<br />

With the segment number can be set.<br />

Before the segment is set a test is run, to check if the entries of the previous segment<br />

contain only valid values. If not, this segment will be initialised and displayed.<br />

If all entries are complete, leave the program 20 with .<br />

Setting program length:<br />

You have for example 20 segments, but only want to run the first three segments you can<br />

set a ‘stop marker’ at the fourth segment. This is done by pressing for the segment<br />

duration and confirmed with .<br />

Stopmarke<br />

or<br />

stop marker<br />

29<br />