unistat control V3.8.1<br />
5. “Redu slave” is activated and the cable is connected – the “master” unit has an<br />
active POKO: the slave thermoregulates with the preset parameters.<br />
The operation modes “Redu slave” and “Redu Master” are alternately displayed on the<br />
second line of the units display. The operation mode POKO master/POKO slave is set in<br />
programme 29 (Redundancy Modes). The “Redu Slave” starts to thermoregulate when the<br />
POKO of the “Redu Master” becomes inactive.<br />
This is the case when:<br />
1. The “Redu Master” is unpowered.<br />
2. The “Redu Master” is not ready yet after starting the equipment.<br />
3. The “Redu Master” goes to error<br />
Behaviour of the “Redu Slave” after activation: Once activated the “Redu slave” can only be<br />
deactivated manually via the keyboard. The mode “Redu Slave” has to be terminated through<br />
programme 29. The thermoregulation continues but can be terminated with the off key.<br />
Effect on the float switch: the same socket is used for both tasks. For this reason, the float<br />
switch is automatically cut off if the “Redu Slave” mode is activated. If the “Redu Slave”<br />
mode is deactivated and the equipment starts with the connected “Redu cable”, an error<br />
message “level error” will be displayed.<br />
Please note: using a second equipment to enhance the security of the installation is a<br />
complex process. The above mentioned functionality is helpful but never the less; it will not<br />
reach the target by itself. For example investigation of the behaviour of the fluid circuit and if<br />
the need of swing-type check valves or the set point on which the slave has to operate and<br />
which procedure will ensure that the slave is really going to be active in an emergency etc…;<br />
still has to be done.<br />
Important note: Please regularly make sure of the correct function of the redundancy<br />
operation before every thermoregulation process.<br />