unistat control V3.8.1<br />

Appendix A - Program Directory<br />

The result is calculated from the following relationship:<br />

- current actual value<br />

= ------------------------------------------<br />

<br />

For which the current actual value is a constant.<br />

For example: The tempering is running at 20°C. Requested is a set ramp of +0.4K/min up to<br />

an end temperature of 73.5°C<br />

20 int 85 max<br />

20.0°C 5 min<br />

enter (or start program 19):<br />

go K/min<br />

??<br />

And then for the raising. The display<br />

indicates:<br />

go K/min<br />

R:0.4 <br />

The ramp is displayed and bottom right is a note of one of the missing entry<br />

values ‘T’ for temperature and ‘t’ for time.<br />

<br />

go K/min<br />

Temp: 73.5<br />

The ramp is started and the following is displayed on the LCD:<br />

22 int 85 max<br />

47.3°C tR 15.5<br />

32,7°C is the given ramp, tR = 15.5 min is the remaining time, which the ramp has still to<br />

run. During this time the gradient of the ramp and the end temperature can be observed if<br />

key is pressed.<br />

go K/min<br />

R: 0.4 T: 73.5<br />

Note: The ramp function is only operable during thermoregulation. The aborting of the ramp is<br />

possible via entering new set points via program 0.<br />

27