unistat control V3.8.1<br />
Appendix A - Program Directory<br />
The result is calculated from the following relationship:<br />
- current actual value<br />
= ------------------------------------------<br />
<br />
For which the current actual value is a constant.<br />
For example: The tempering is running at 20°C. Requested is a set ramp of +0.4K/min up to<br />
an end temperature of 73.5°C<br />
20 int 85 max<br />
20.0°C 5 min<br />
enter (or start program 19):<br />
go K/min<br />
??<br />
And then for the raising. The display<br />
indicates:<br />
go K/min<br />
R:0.4 <br />
The ramp is displayed and bottom right is a note of one of the missing entry<br />
values ‘T’ for temperature and ‘t’ for time.<br />
<br />
go K/min<br />
Temp: 73.5<br />
The ramp is started and the following is displayed on the LCD:<br />
22 int 85 max<br />
47.3°C tR 15.5<br />
32,7°C is the given ramp, tR = 15.5 min is the remaining time, which the ramp has still to<br />
run. During this time the gradient of the ramp and the end temperature can be observed if<br />
key is pressed.<br />
go K/min<br />
R: 0.4 T: 73.5<br />
Note: The ramp function is only operable during thermoregulation. The aborting of the ramp is<br />
possible via entering new set points via program 0.<br />
27