Appendix A -- Program Directory<br />

unistat control V3.8.1<br />

Appendix A - Program Directory<br />

The setting of operating modes or unit parameters is carried out using programs. These are<br />

stored in an extensive memory in the control panel. Some programs only switch from one<br />

operating mode to an alternative (for example ‘control mode’ program 3), with others<br />

parameters have to be entered (for example ‘set point’ program 0) and some need a proper<br />

dialogue (for example ‘setting heating program’ program 20).<br />

The program selection is initiated by pressing , after that the program number must be<br />

entered, then press . A short description of the program appears on the LCD display<br />

unit together with a display of the current setting.<br />

Entering the numerical values for the parameters, for example the set point, is done via the<br />

number keys … , sometimes also by using the decimal point and the minus<br />

sign . The programming is confirmed by pressing .<br />

Corrections are possible by pressing . A programming mistake is stopped by pressing<br />

. Should a mistake be confirmed with , it can still be rectified by calling up the<br />

program and entering the new parameter.<br />

! Not every control panel is equipped with all programs! This depends on the software<br />

status!<br />

The software of the unistat control monitors logic discrepancies. A change of the minimum –<br />

maximum set point, for example, during activated circulation is not possible. If you try such<br />

change you will be warned (display ‘ Not allowed!’). The programmed change is rejected.<br />

Program 0: Setting set point<br />

After selection, ‘enter set point’ appears on LCD display. The set point is entered with <br />

to , and keys. The actual programming is done by pressing . The<br />

value of the set point is depending on the currently set minimum set point and maximum set<br />

point. The rule is:<br />

Minimal set point < = set point < = maximum set point.<br />

If this rule is not obeyed, an error message appears and the setting is ignored. The step size<br />

for the entry is 0.1K. You can choose to enter the second digit after the comma, by not<br />

pressing at the end, but by just entering another digit. With this you can reach a<br />

setting resolution of 0.01K.<br />

entering set point of 20.5 °C,<br />

(without !) setting a set<br />

point of 20.55°C.<br />

The activation of this program is achieved by pressing , this is the<br />

same as pressing the Hotkey .<br />

Program 1: Setting Of Minimum Set Point<br />

After selection, the program name appears and the current minimum set point appears.<br />

The value is entered by pressing keys … and . Only whole numbers are<br />

possible, no digits after comma. The entering is done by pressing . The value of the<br />

minimum set point is depending on the current set point and the unit parameters<br />

(Lowest permissible temperature). The rule is:<br />

Lowest permissible temperature < = minimum set point < = set point.<br />

If this rule is broken, an error message appears, and the setting is ignored.<br />

19