R a r e E a r t h E l e m e n t E x p l o r a t i o n P r o g r a m M a r c h - 2 0 1 3<br />

W i n t e r - S u m m e r 2 0 1 2 P a g e | 72<br />

Item 15. SAMPLE PREPARATION, ANALYSES AND SECURITY<br />

15.1 SAMPLE PREPARATION<br />

Samples were sent to SGS Minerals Services (“SGS”) of Lakefield in Ontario, where all<br />

the samples were prepared (crushed, ground, dried) and analyzed. A summary of the<br />

sample preparation is following but the detailed SGS analysis and preparation techniques<br />

can be consulted on their website (www.sgs.com).<br />

As a routine practice with core, the entire sample is crushed to a nominal minus 10 mesh<br />

(2 mm), mechanically split via a riffle splitter in order to divide the sample into a 250 gr<br />

sub-sample for analysis and the remainder is stored as a reject. Samples are pulverized to<br />

85% passing 75 micron (200 mesh) or otherwise specified by client. SGS used their own<br />

sample preparation control for more accuracy (Table 9).<br />

Table 9: Internal quality control for sample preparation by SGS Ontario.<br />

Crushing<br />

Parametres<br />

Frequency<br />

Quality Control Requirement<br />

Prep. Blank At the start of batch To Clean Crusher<br />

Prep. Replicates every 50 samples 75% passing 10 mesh (2mm)<br />

Passing Checks Every 50 samples 75% passing 10 mesh (2mm)<br />

Regarding samples analysis, SGS laboratory used the following methods:<br />

• ICM90A, for 55 elements, by sodium peroxide fusion and a combination of<br />

Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES), and<br />

Inductively Coupled Plasma Mass Spectrometry (ICP-MS).<br />

• IMS91B by sodium peroxide fusion / ICP-MS, for lanthanides surplus upper<br />

limit.<br />

15.2 ANALYSIS<br />

15.2.1 ICM90A<br />

Crushed and pulverized rocks are fused by Sodium peroxide in graphite crucibles and<br />

dissolved using diluted HNO3. During digestion the sample is split into 2 and half is<br />

given to ICP-OES and the other half is given to ICP¬MS. The digested sample solution<br />

is analyzed by ICP-OES and ICP-MS. Samples are analyzed against known calibration<br />

materials to provide quantitative analysis of the original sample.<br />

The data results fed to the SGS Laboratory Information Management System (SLIM)<br />

with secure audit trait are exported online via computer. This method has been fully<br />

validated for the range of samples typically analyzed. Method validation includes the use<br />

of certified reference materials, replicates and blanks to calculate accuracy, precision,<br />

linearity, range, detection limits, and limit of quantification, specificity and measurement<br />

uncertainty.<br />