R a r e E a r t h E l e m e n t E x p l o r a t i o n P r o g r a m M a r c h - 2 0 1 3<br />
W i n t e r - S u m m e r 2 0 1 2 P a g e | 73<br />
Table 10: Elements analyzed by ICM 90A.<br />
Element<br />
Reporting<br />
Limit<br />
(ppm)<br />
Upper<br />
Limit<br />
Element<br />
Reporting<br />
Limit<br />
(ppm)<br />
Upper<br />
Limit<br />
Element<br />
Reporting<br />
Limit<br />
(ppm)<br />
Upper<br />
limit<br />
Element<br />
Reporting<br />
Limit<br />
(ppm)<br />
Upper<br />
Limit<br />
Ag 1.00 0.01% Er 0.05 0.10% Mn 10 10% Tb 0.05 0.10%<br />
Al 0.01(%) 25% Eu 0.05 0.10% Mo 2.00 1.0% Th 0.10 0.10%<br />
As 5.00 10% Fe 0.01(%) 30% Nb 1.00 1.0% Ti 0.01(%) 25%<br />
Ba 0.50 1.0% Ga 1.00 0.10% Nd 0.10 1.0% TI 0.50 0.10%<br />
Be 5.00 0.25% Gd 0.05 0.10% Ni 5.00 1.0% Tm 0.05 0.10%<br />
Bi 0.10 0.10% Ge 1.00 0.10% P 0.01(%) 25% Ta 0.50 1.0%<br />
Ca 0.01(%) 35% Hf 1.00 1.0% Pb 5.00 1.0% U 0.05 0.1%<br />
Cd 0.20 1.0% Ho 0.05 0.10% Pr 0.05 0.1% V 5.00 1.0%<br />
Ce 0.10 1.0% In 0.20 0.10% Rb 0.20 1.0% W 1.00 1.0%<br />
Co 0.50 1.0% K 0.01(%) 25% Sb 0.50 1.0% Y 0.50 0.1%<br />
Cr 10 10% La 0.10 1.0% Sc 5.00 5.0% Yb 0.10 0.1%<br />
Cs 0.10 1.0% Li 10 5.0% Sm 0.10 0.1% Zn 5.00 1.0%<br />
Cu 5.00 1.0% Lu 0.05 0.10% Sn 1.00 1.0% Zr 0.50 1.0%<br />
Dy 0.05 0.1% Mg 0.01(%) 30% Sr 0.10 1.0%<br />
15.2.2 ICPMS<br />
ICPMS diluted samples to be analyzed on Elan 9000 for 90A samples and Nexion for<br />
91B packages.<br />
Working Calibration solutions and 2 nd source calibration check solution was prepared for<br />
each analysis run. Re-calibration was done before the analysis of each tray. Additional<br />
fusion QC was analyzed every other tray in addition to the QC on each tray.<br />
REE interference corrections were evaluated and corrected.<br />
15.2.3 ICP/OES<br />
Samples are analyzed with a minimum of 10 certified reference materials for the required<br />
analyses, all prepared by sodium peroxide fusion. Every 10 th sample is prepared and<br />
analyzed in duplicate; a blank is prepared every 30 samples and analyzed. Samples are<br />
analyzed using a Varian 735ES ICP or a Thermo 6500 ICAP and the method of internal<br />
standardization.<br />
For High concentration of REE, the IMS91B analysis technic was used.<br />
Crushed and pulverized rock, samples (0.20 gr) are fused by Sodium peroxide in glassy<br />
carbon crucibles in a muffle furnace and dissolved using diluted HNO3. The fused<br />
solution sample is aspirated into the Inductively Coupled Plasma Dynamic Reaction Cell<br />
Mass Spectrometre (ICP-DRC-MS). Samples are analyzed against known calibration<br />
materials to provide quantitative analysis of the original sample.<br />