Installation<br />
Figure 4 Isokinetic probe with tubing and zero count filter<br />
1 Hose barb fitting 3 Tubing<br />
2 Isokinetic probe 4 <strong>MET</strong> <strong>ONE</strong> <strong>HHPC</strong>–2<br />
3.3.3 Install the zero count filter<br />
In a clean-room application, the zero count filter (refer to Figure 3 on page 11, item 2)<br />
verifies if the <strong>MET</strong> <strong>ONE</strong> <strong>HHPC</strong>–2 is not counting electrical noise signals from internal<br />
components and is not subject to external interference. In indoor air quality and other<br />
applications, the zero count filter cleans out the sensor immediately after a high<br />
concentration sampling.<br />
To connect the zero count filter:<br />
1. Remove the isokinetic probe from the <strong>MET</strong> <strong>ONE</strong> <strong>HHPC</strong>–2.<br />
2. Attach the intake hose barb (MP000026–01).<br />
3. Secure the high purity tubing (VP792002) to the hose barb.<br />
4. Secure the opposite end of the tubing to the zero count filter (VP212808).<br />
Note: Make sure that the arrow of airflow direction or the indicator on the zero count filter points<br />
towards the tubing.<br />
13