<strong>FUJI</strong> <strong>HART</strong> <strong>EXPLORER</strong>Dactrim functionIntroductionThis function will adjust the output signal. It willTRIM THE ZERO OR 4 MILLIAMP POINT OF THE DIGITAL TO ANALOG CONVERTER SOTHAT THE CONNECTED CURRENT METER READS 4 MILLIAMP.TRIM THE GAIN OR 20 MILLIAMP POINT OF THE DIGITAL TO ANALOG CONVERTER SOTHAT THE CONNECTED CURRENT METER READS 20 MILLIAMP.Procedure14/09/2010 41/76