<strong>FUJI</strong> <strong>HART</strong> <strong>EXPLORER</strong>When clicking on START button, the following window is displayed.Enter the output signal displayed on the milliamp – meter connected to the transmitter in the “actualcurrent” space.- first for the LRV- next for the URVCalibrate the output signal only with a high accurate milliamp – meter (3 digits after the point)Close the window on “Close” button.14/09/2010 65/76