Control Commands7.2i7.2.1ModesThe sign ESC (1Bh) finishes the data output. Now the sensor waits for a new command.DT......DistancetrackingEssential input parameters: SA, SD, SE, SF, ST, OFEffect on:RS232/RS422, digital switching output, analog output-- Distance measurement at different kinds of surfaces (varying reflectance).-- Permanent evaluation of the sensitive laser radiation.• changing reflectance: longer measuring time• sudden jumps in distance: longer measuring timeThe minimum time to measure is 160 ms, the maximum time is 6 s. If the measuring signal fails to reach aspecified quality within six seconds, an error message is output.The time to measure may also be limited by setting the ST parameter to a desired value.7.2.2 DS ...... 7 m Distance TrackingEssential input parameters:Effect on:SA, SD, SE, SF, ST, OFRS232/RS422, digital switching output, analog output-- Distance measurement at different kinds of surfaces at close range up to 7 m-- Higher measurement rate compared to DT measuring mode.-- Within the range from 0.1 m to 0.5 m, measuring accuracy is restricted.Measuring time (time to measure) can be limited via ST parameter settings.<strong>optoNCDT</strong> <strong>ILR</strong> <strong>1181</strong> / <strong>1182</strong>Page 28