12 Appendix12.1 Updated SpecificationsElectrical ParametersPower SourceBatteryDisplayNumber of CharactersHeightWidthIllumination<strong>Interface</strong>s120 V A/C (Home power)When unplugged from wall device switches toRhino 12 Volt 3.4Ah Sealed Lead Acid Battery20 min.3.5” min.6” min.High contrast, easily visible in strongsunlightMust effectively operate infusion pump viainterface.Buttons 1.5” x 1.5”Actuation pressureMust resist inadvertent activation, but remaineasy to operateSize 11” x 5.5” x 3.25”Weight5 lbs max. for interface enclosure, 10 lbs max forbattery/pumping enclosureDurabilityOperate nominally after 1’ dropVibrationOperate nominally after 2Hz sinusoidEnvironmentalLocationDustOperating TempStorage TempHome and sterile hospital environmentsSealed container will resist dust. Polypropylenefilter under possible speaker cutouts32 F to 105 F0 F to 120 FSoftwareMachine DependentMust interface with infusion pump to controlpump operation via pumping assembly fromBaxter Flo-Gard series of infusion pumps112