2.2.2.2 Toggle Switch / JoystickReplacing the scrolling arrow buttons with a single axis joystick will providepatients with an easier means to choosing the correct menu or number from theLCD. This particular joystick comes with a restrictor plate to make it one axis fromits normal two. Reducing the joystick to one axis of motion allows only up ordown inputs from the patient. The J1 low profile joystick as seen in Fig. 2.2.2.2.1on the next page has a plus or minus twelve degree movement from centerwhich is enough to avoid inadvertent inputs. This degree of movement also isn’tlarge enough to make the motion strenuous to the user when scrolling throughmenus or numbers for selection.Figure 2.2.2.2.1: Picture of J1 Low Profile "Switch Stick" Joystick from etisystems.com29