EKSpot MALDI Spotter System Operator Guide - Eksigent
EKSpot MALDI Spotter System Operator Guide - Eksigent
EKSpot MALDI Spotter System Operator Guide - Eksigent
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
InstallationInstall the <strong>EKSpot</strong> <strong>MALDI</strong> <strong>Spotter</strong> <strong>System</strong>1. Place the spotter to the right of the NanoLC system.Figure 3-1Position of the <strong>EKSpot</strong> <strong>MALDI</strong> spotter and the NanoLC system2. Assemble the spotter and place it on the base plate.Refer to the PAL manual for more detailed instructions.3. Secure the spotter to the base plate by tightening the two set screws on each side ofthe base plate, if necessary.Note: The set screws are for alignment purposes only. They do notsecure the spotter to the base plate.4. Install the stack holder to the right of the spotter as shown in Figure 3-1.5. Connect the spotter to the computer.Use COM port 1 to minimize communication errors.<strong>EKSpot</strong> <strong>MALDI</strong> <strong>Spotter</strong> <strong>System</strong><strong>Operator</strong> <strong>Guide</strong>12 of 38 September 2012