Infrared Gas AnalyzerInstallation• Solenoid Valve—for introducing calibration gas andOFF/ON flow of sample gas.• Membrane Filter—PTFE filter for rempval of fine dustparticles.• Flowmeter—adjusts and monitors the sample gas flowrate.• Standard Gas—a bank of calibration gases used for settingzero and span of the analyzer for each species monitored.• Zirconia Sensor— an external zirconia sensor is used formeasuring the oxygen concentration in the sample gas. Someinstruments use a built-in paramagnetic oxygen sensorinstead aof an external zirconia sensor.• NO 2 /NO Converter—added to the NO analysis circuit whichuses an efficient catalyst for converting NO 2 to NO foranalysis.Figure 2-8: Five-Component Analysis System<strong>Teledyne</strong> <strong>Analytical</strong> <strong>Instruments</strong> 31