13.07.2015 Views

Model 7600 - Teledyne Analytical Instruments

Model 7600 - Teledyne Analytical Instruments

Model 7600 - Teledyne Analytical Instruments

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Infrared Gas AnalyzerInstallation• Solenoid Valve—for introducing calibration gas andOFF/ON flow of sample gas.• Membrane Filter—PTFE filter for rempval of fine dustparticles.• Flowmeter—adjusts and monitors the sample gas flowrate.• Standard Gas—a bank of calibration gases used for settingzero and span of the analyzer for each species monitored.• Zirconia Sensor— an external zirconia sensor is used formeasuring the oxygen concentration in the sample gas. Someinstruments use a built-in paramagnetic oxygen sensorinstead aof an external zirconia sensor.• NO 2 /NO Converter—added to the NO analysis circuit whichuses an efficient catalyst for converting NO 2 to NO foranalysis.Figure 2-8: Five-Component Analysis System<strong>Teledyne</strong> <strong>Analytical</strong> <strong>Instruments</strong> 31

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!