13.07.2015 Views

Model 7600 - Teledyne Analytical Instruments

Model 7600 - Teledyne Analytical Instruments

Model 7600 - Teledyne Analytical Instruments

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Infrared Gas AnalyzerOperationThe numbered call outs in Figure 3-4 refer to:No. Name Function(1) ComponentdisplayDisplays component of instantaneousvalue, corrected instantaneous value,corrected average value, etc.(2) ConcentrationdisplayDisplays measured value of concentration.(3) Range display Displays range values.(4) Unit display Displays unit with ppm and vol %.(5) Peak alarmcomponent(6) Peak alarmconcentration(7) Peak alarmtimes(8) Peak alarmunit displayDisplays peak alarm component.Displays peak alarm concentration display.(Upper limit value)Displays the alarm times exceeding thepeak value.Displays units of peak alarm with times/H.Instantaneous value and concentration value:The concentration display of Ch (component) where samplingcomponents such as “CO 2 ”, “CO” or “O 2 are displayed in the componentdisplay, indicates current concentration values of the measuredcomponents contained in gas that is now under measurement.O 2 correction concentration values:Ch components where “cv**” is displayed as “cv CO” in thecomponent display are calculated values. They are obtained from thefollowing equation by setting sampling components, O 2 instantaneous/concentration values and O 2 correction reference value (see item 6.8).⎛ 21- On ⎞Correction output = ⎜ ⎟ × Cs⎝ 21−Os ⎠<strong>Teledyne</strong> <strong>Analytical</strong> <strong>Instruments</strong> 39

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!