Infrared Gas AnalyzerOperation4. In the “Calibration Parameters” screen that appears, pointthe cursor to “Calibration Value” by pressing the UP orDOWN key. Press the ENT key.To change the concentration of a calibration gas:1. Point the cursor to the channel you want to set by usingthe UP or DOWN key. Press the ENT key.2. Select the concentration item you want to set by pressingthe UP or DOWN or SIDE key. Press the ENT key, andthe selected value is highlighted.3. Enter the calibration gas concentration values (zero andspan). For any numerical value entry, press the UP orDOWN key to increase or decrease a single digit. Selectthe next digit by pressing the SIDE key.4. When the concentration is correct, save the entry bypressing the ENT key. The saved value becomes validfrom the next calibration process.<strong>Teledyne</strong> <strong>Analytical</strong> <strong>Instruments</strong> 47