<strong>Retriever</strong> <strong>IV</strong> <strong>Fraction</strong> <strong>Collector</strong>Section 4 Theory of OperationWhen the last position on the red stop rack is reached, the outputof U127D will go low resetting the motor control flip-flop U114Bto stop the motor. In addition, the output of U115A will set theend of run flip-flop U128A. When U128A is set the Q output goeslow preventing any additional input counts on U105, also turningthe security valve output off, the pump stop output of the countconnector on, and operating the pump isolation relay K101. Thisresults in turning off any pumps connected to the fraction collectorand closing the security valve.4.1.1 Timing Diagram To illustrate the timing and relationship of signals during operation,selected waveforms are shown on the timing diagram,Figure 4-2. The timing diagram assumes the front panel switchesare set as shown below.TIME/COUNT - set at 001X1 /X10 - set XlTIME/VOL/DROP - set at TIMEDELAY/TIME - set at 004Figure 4-2 Timing Diagram4-5