16 Unpacking and Setup<br />

Connection Function<br />

5TTL TTL-compatible output<br />

Levels:<br />

� passive 0 V<br />

� active 5 V<br />

Pulse:<br />

� 5 V for at least 1000 ms<br />

6OC TTL output<br />

Levels:<br />

� passive 0 V<br />

� active 5 V<br />

Pulse:<br />

� 5 V for at least 1000 ms<br />

GND Reference point of the voltage at the signal<br />

inputs.<br />

Relay contact<br />

The contact is on a floating basis. Its setting<br />

depends on the settings in the Control Unit<br />

or software.<br />

Steady-rate signal:<br />

� passive = open relay contact<br />

� active = closed relay contact<br />

Pulse:<br />

� Closed relay contact for at least<br />

1000 ms<br />

Permissible load of the relay contact: 1 A/<br />

24 V DC<br />

Relay contact<br />

The contact is on a floating basis. Its setting<br />

depends on the settings in the Control Unit<br />

or software.<br />

Steady-rate signal:<br />

� passive = open relay contact<br />

� active = closed relay contact<br />

Pulse:<br />

� Closed relay contact for at least<br />

1000 ms<br />

Permissible load of the relay contact: 1 A/<br />

24 V DC<br />

Analog GND Reference point of the voltage at the signal<br />

inputs.<br />

Analog out Voltage range 0 – 5 V, scalable<br />

Analog in Voltage range 0 – 10 V<br />

10 V according to maximum flow rate