16 Unpacking and Setup<br />
Connection Function<br />
5TTL TTL-compatible output<br />
Levels:<br />
� passive 0 V<br />
� active 5 V<br />
Pulse:<br />
� 5 V for at least 1000 ms<br />
6OC TTL output<br />
Levels:<br />
� passive 0 V<br />
� active 5 V<br />
Pulse:<br />
� 5 V for at least 1000 ms<br />
GND Reference point of the voltage at the signal<br />
inputs.<br />
Relay contact<br />
The contact is on a floating basis. Its setting<br />
depends on the settings in the Control Unit<br />
or software.<br />
Steady-rate signal:<br />
� passive = open relay contact<br />
� active = closed relay contact<br />
Pulse:<br />
� Closed relay contact for at least<br />
1000 ms<br />
Permissible load of the relay contact: 1 A/<br />
24 V DC<br />
Relay contact<br />
The contact is on a floating basis. Its setting<br />
depends on the settings in the Control Unit<br />
or software.<br />
Steady-rate signal:<br />
� passive = open relay contact<br />
� active = closed relay contact<br />
Pulse:<br />
� Closed relay contact for at least<br />
1000 ms<br />
Permissible load of the relay contact: 1 A/<br />
24 V DC<br />
Analog GND Reference point of the voltage at the signal<br />
inputs.<br />
Analog out Voltage range 0 – 5 V, scalable<br />
Analog in Voltage range 0 – 10 V<br />
10 V according to maximum flow rate