44 Maintenance and Care<br />
Attaching capillaries Process Figure<br />
3. Using an open-end<br />
wrench, hold the inlet<br />
fitting 5 in place.<br />
4. Tighten the fitting 4 at<br />
the pump head.<br />
5. If the capillary cannot<br />
be screwed tight, then<br />
use a new clamping<br />
ring!<br />
Maintaining the Check Valves<br />
Dirty check valves do not properly open and close. They cause<br />
pressure fluctuations and irregular flow. If it is impossible to<br />
clean the check valves, replace the whole unit.<br />
Prerequisite � <strong>Pump</strong> head has been flushed.<br />
� Capillaries are disconnected.<br />
� <strong>Pump</strong> head has been dismounted.<br />
Procedure � Unscrew the outlet and inlet fittings.<br />
� Remove the check valve.<br />
� Insert the check valve.<br />
� Screw in the outlet and inlet fittings.<br />
Tools � Open-end wrench, sizes 10, 17<br />
Removing the Check Valves<br />
Process Figure<br />
6. Unscrew the outlet fitting<br />
1.<br />
7. Remove the check valve<br />
2.<br />
8. Unscrew the inlet fitting<br />
3.<br />
9. Remove the check<br />
valve.<br />
Intermediate result Check valves are removed.<br />
Next steps Replace or clean the check valves.<br />
Fig. 40 <strong>Pump</strong> head<br />
1<br />
2<br />
3<br />
4<br />
5<br />
Fig. 41 Removing the check<br />
valve