44 Maintenance and Care<br />

Attaching capillaries Process Figure<br />

3. Using an open-end<br />

wrench, hold the inlet<br />

fitting 5 in place.<br />

4. Tighten the fitting 4 at<br />

the pump head.<br />

5. If the capillary cannot<br />

be screwed tight, then<br />

use a new clamping<br />

ring!<br />

Maintaining the Check Valves<br />

Dirty check valves do not properly open and close. They cause<br />

pressure fluctuations and irregular flow. If it is impossible to<br />

clean the check valves, replace the whole unit.<br />

Prerequisite � <strong>Pump</strong> head has been flushed.<br />

� Capillaries are disconnected.<br />

� <strong>Pump</strong> head has been dismounted.<br />

Procedure � Unscrew the outlet and inlet fittings.<br />

� Remove the check valve.<br />

� Insert the check valve.<br />

� Screw in the outlet and inlet fittings.<br />

Tools � Open-end wrench, sizes 10, 17<br />

Removing the Check Valves<br />

Process Figure<br />

6. Unscrew the outlet fitting<br />

1.<br />

7. Remove the check valve<br />

2.<br />

8. Unscrew the inlet fitting<br />

3.<br />

9. Remove the check<br />

valve.<br />

Intermediate result Check valves are removed.<br />

Next steps Replace or clean the check valves.<br />

Fig. 40 <strong>Pump</strong> head<br />

1<br />

2<br />

3<br />

4<br />

5<br />

Fig. 41 Removing the check<br />

valve