16.02.2013 Views

AZURA Pump P 2.1L user manual

AZURA Pump P 2.1L user manual

AZURA Pump P 2.1L user manual

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Legend<br />

1 Capillary screw fitting<br />

2 Allen screws<br />

3 Outlet fittings<br />

4 Inlet fittings<br />

5 Eluent inlet<br />

42 Maintenance and Care<br />

Screw Fittings on the <strong>Pump</strong> Head<br />

1<br />

Leaks in the Capillary Screw Fittings<br />

CAUTION! Damage to the device possible! If leaks occur<br />

after maintenance and assembly, replace the<br />

capillary connections with new ones!<br />

Replacing the <strong>Pump</strong> Head<br />

Depending on the requirements of the <strong>user</strong>, different pump<br />

heads are used. The chapter accessories and spare parts contains<br />

a pump head overview.<br />

Prerequisite � The pump head has been purged.<br />

� Hoses are disconnected.<br />

Procedure � Loosen the capillary fittings.<br />

� Loosen the Allen screws on the pump head.<br />

� Remove the pump head.<br />

Tools � Allen wrench size 3<br />

� Open-end wrench, sizes 10, 17<br />

2<br />

3<br />

4<br />

5<br />

Fig. 37 Screw fittings on the pump head<br />

Dismounting the <strong>Pump</strong> Head<br />

WARNING! Aggressive or toxic solvent residue can irritate<br />

the skin!<br />

Wear protective gloves.<br />

Flush the pump head before exchanging it.<br />

CAUTION! Damage to the pump pistons due to jamming of<br />

the pump head.<br />

Loosen diagonally opposite fastening screws<br />

evenly by one turn.