54 Troubleshooting<br />
System message Solution<br />
Maximum pressure!<br />
System stopped<br />
Minimum pressure!<br />
System stopped<br />
Motor failure: max<br />
current<br />
N No gradient is available<br />
in isocratic<br />
mode<br />
� Reduce the pressure or adjust the<br />
upper pressure limit.<br />
� Restart the system<br />
� Increase the pressure or adjust the<br />
lower pressure limit.<br />
� Restart the system<br />
Switch the device off and on. Inform<br />
the technical support of the manufacturer<br />
in case the system message<br />
repeats itself.<br />
Change the setup settings or change<br />
the gradient in the program or in<br />
setup.<br />
No link available Create a link and edit it.<br />
No link available. Pls<br />
edit link first<br />
No time table to<br />
start<br />
Non-existing component<br />
is set to non-<br />
0 value<br />
P Program does not<br />
exist<br />
Create a link and edit it.<br />
Edit the data by means of the chromatography<br />
software.<br />
Switch on the channel or edit the data<br />
using the chromatography software.<br />
Create a program.<br />
Program is running Quit program or wait until program<br />
has been completed.<br />
Program not compatible<br />
with pump<br />
head<br />
S Sum of components<br />
is not 100<br />
T This link is used in<br />
WAKEUP<br />
This program is used<br />
in a link<br />
This program is used<br />
in WAKEUP<br />
Modify the program or replace the<br />
pump head.<br />
Change the entry.<br />
First quit or delete wakeup program<br />
(wu = Wake Up), then edit or delete<br />
link.<br />
First pause or delete the link, then edit<br />
or delete data by means of the chromatography<br />
software.<br />
First quit or delete wakeup program<br />
(wu = Wake Up), then edit or delete<br />
data by means of the chromatography<br />
software.<br />
Time already exists Correct the time entry.