20 Initial Startup<br />
Connection of the Eluent Line<br />
Prerequisite The eluent inlet is connected.<br />
Procedure Connect the eluent line to the pump head.<br />
Note If connecting a pump head from 100 – 500 ml, make sure that<br />
the tapered side of the cutting ring is pointed towards the fastening<br />
screw of the Teflon tube.<br />
Connecting the eluent<br />
line<br />
<strong>Pump</strong> head:<br />
� 100 ml<br />
� 250 ml<br />
� 500 ml<br />
Connecting the eluent<br />
line<br />
<strong>Pump</strong> head:<br />
1000 ml<br />
Process Figure<br />
1. Push the Teflon hose 1<br />
through the fastening<br />
screw 2 and the cut-<br />
12 3 4<br />
ting ring 3.<br />
2. Insert the hose end as<br />
far as possible into the<br />
eluent inlet fitting 4 of<br />
the pump head.<br />
3. Tighten the fastening<br />
screw by hand. Fig. 9 Connecting the eluent<br />
line to the pump head<br />
Process Figure<br />
1. Push the Teflon hose<br />
directly on the olivetype<br />
tube fitting 1.<br />
2. Fasten the Teflon hose<br />
with hose clamp.<br />
1<br />
Result The eluent line is connected.<br />
Next steps Check the seal and lines of the connections.<br />
Fig. 10 Eluent line and 1000 ml<br />
pump head