36 Initial Startup<br />
Switching On the <strong>Pump</strong><br />
CAUTION! Damage to the pump head in case it runs dry.<br />
Ensure that liquids runs through pump head and<br />
piston backflushing.<br />
Prerequisite � Liquid container is sufficiently filled.<br />
� Piston backflushing is connected.<br />
� Washing container is sufficiently filled.<br />
Switching on the<br />
pump<br />
Process Figure<br />
1. Switch on the pump at<br />
the power switch 1 on<br />
the rear side.<br />
2. Wait until the pump<br />
has completed the selftest.<br />
3. If the self-test has been<br />
successfully completed,<br />
the LED on the<br />
right lights up green.<br />
Result The pump is now ready for operation. If the test fails an error<br />
message will be displayed. Contact the Technical support of the<br />
manufacturer if the error occurs several times.<br />
Flushing the <strong>Pump</strong><br />
Fig. 35 Power switch<br />
To flush the pump, insert the inlet hoses into the storage containers<br />
and start the pump with an intermediate flow rate. As<br />
the pump is self-priming, the venting screw can be open.<br />
CAUTION! Damage to the columnn due to venting!<br />
Open the venting screw or remove column!<br />
When is flushing Flush the pump in the following cases:<br />
required?<br />
� At initial startup to eliminate air bubbles in hoses and capillaries.<br />
� When changing solvents.<br />
� After using buffer solutions to eliminate salt residues.<br />
Note<br />
Use the solvent for flushing that is to be used in the subsequent<br />
application.<br />
If you used a buffer solution, pay attention to choosing a solvent<br />
for flushing in which the buffer solution is soluble.<br />
The purging process of the pump is limited to a maximum pressure<br />
of 5 MPa. If this value is exceeded during the purging process,<br />
the pump switches off automatically. If you are using very<br />
small hoses and capillaries, the pressure can be too high.<br />
1