36 Initial Startup<br />

Switching On the <strong>Pump</strong><br />

CAUTION! Damage to the pump head in case it runs dry.<br />

Ensure that liquids runs through pump head and<br />

piston backflushing.<br />

Prerequisite � Liquid container is sufficiently filled.<br />

� Piston backflushing is connected.<br />

� Washing container is sufficiently filled.<br />

Switching on the<br />

pump<br />

Process Figure<br />

1. Switch on the pump at<br />

the power switch 1 on<br />

the rear side.<br />

2. Wait until the pump<br />

has completed the selftest.<br />

3. If the self-test has been<br />

successfully completed,<br />

the LED on the<br />

right lights up green.<br />

Result The pump is now ready for operation. If the test fails an error<br />

message will be displayed. Contact the Technical support of the<br />

manufacturer if the error occurs several times.<br />

Flushing the <strong>Pump</strong><br />

Fig. 35 Power switch<br />

To flush the pump, insert the inlet hoses into the storage containers<br />

and start the pump with an intermediate flow rate. As<br />

the pump is self-priming, the venting screw can be open.<br />

CAUTION! Damage to the columnn due to venting!<br />

Open the venting screw or remove column!<br />

When is flushing Flush the pump in the following cases:<br />

required?<br />

� At initial startup to eliminate air bubbles in hoses and capillaries.<br />

� When changing solvents.<br />

� After using buffer solutions to eliminate salt residues.<br />

Note<br />

Use the solvent for flushing that is to be used in the subsequent<br />

application.<br />

If you used a buffer solution, pay attention to choosing a solvent<br />

for flushing in which the buffer solution is soluble.<br />

The purging process of the pump is limited to a maximum pressure<br />

of 5 MPa. If this value is exceeded during the purging process,<br />

the pump switches off automatically. If you are using very<br />

small hoses and capillaries, the pressure can be too high.<br />