PERFORMANCES OF THE ASSAY<br />
SPECIFICITY<br />
No cross-rea<strong>ct</strong>ions were observed with the present method.<br />
SENSITIVITY<br />
The sensitivity was calculated based upon the calibration curve and expressed as the minimal dose<br />
showing a significant difference from the Zero Calibrator (mean value + 3 S.D.). This dose is 1.5 ng/mL.<br />
PRECISION<br />
Precision was evaluated upon intra- and inter-assay variability, in 3 sera at different Ferritin<br />
concentrations.<br />
Intra-assay (Repeatability)<br />
Serum Mean ± S.D. C.V. Replicates<br />
(ng/mL) % no.<br />
a 32.6 ± 1.4 4.2 10<br />
b 144.7 ± 7.3 5.0 10<br />
c 373.7 ± 20.2 5.4 10<br />
Inter-assay (Reproducibility)<br />
Serum Mean ± S.D. C.V. Assays<br />
(ng/mL) % no.<br />
a 27.6 ± 2.05 7.4 10<br />
b 160.5 ± 8.8 5.4 10<br />
c 1120 ± 91.0 8.1 10<br />
ACCURACY<br />
Accuracy of the method was checked by recovery and parallelism tests:<br />
Recovery Test<br />
Known amounts of Ferritin have been added to a normal serum and tested.<br />
Added Expe<strong>ct</strong>ed Measured Recovery<br />
(ng/mL) (ng/mL) (ng/mL) %<br />
S1 ---- 10 ----<br />
S1 + 1000 1010 1025 101<br />
S1 + 500 510 540 105<br />
S1 + 250 260 245 94<br />
S1 + 50 60 58 96<br />
S1 + 20 30 31 103<br />
Parallelism Test<br />
A high Ferritin concentration serum was tested at different dilutions with the Zero Calibrator.<br />
Dilution Expe<strong>ct</strong>ed Measured<br />
(ng/mL) (ng/mL)<br />
S2 undiluted ---- 1493<br />
1:2 746.5 730<br />
1:4 373.2 421<br />
1:8 186.6 181<br />
1:16 93.3 101<br />
1:32 46.6 45<br />
1:64 23.3 24.7<br />
1:128 11.6 12.5<br />
KP33CT – FERRITINA IRMA<br />
M99 – Rev.09 – 12/2004 – Pag. 11/32