18.05.2013 Views

PROCEEDINGS OF THE 7 INTERNATIONAL ... - Fizika

PROCEEDINGS OF THE 7 INTERNATIONAL ... - Fizika

PROCEEDINGS OF THE 7 INTERNATIONAL ... - Fizika

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

C.-M. Nilsson / Medical Physics in the Baltic States 7 (2009) 60 - 63<br />

together with another medical physicist to minimize the<br />

risks for mistakes.<br />

Fig. 3. In house built water phantom for measurements of<br />

absolute dose. Room for an ion-chamber is also seen.<br />

In the larger annual services that lasts a week, a large<br />

water tank (see Fig. 4) is used to take up dose profiles<br />

and depth doses. These are compared with the results of<br />

previous measurements of absorbed dose in the water<br />

tank and also the profiles taken up previously.<br />

Depth dose and profiles for open and wedge fields,<br />

absolute dose, TMR 20 10, and R50 are measured. These<br />

services also include preventive exchanges on the<br />

accelerators for example if one sees a MLC leaf<br />

beginning to falter, one exchanges it beforehand, or<br />

lubricates etc, all to keep the accelerator working<br />

properly.<br />

Fig. 4. Water tank from IBA, where an electrometer, diode,<br />

computer etc. is connected.<br />

Patient Specific QC<br />

IMRT QC<br />

QC of IMRT is performed before each patient starts<br />

his/her treatment. The plan is exported and calculated<br />

on a phantom, IBA MatriXX (see Fig. 5).<br />

62<br />

Fig. 5. IMRT phantom from IBA, with MULTICube<br />

polystyrene.<br />

The phantom is then irradiated in two ways. One series<br />

of irradiations where the phantom is irradiated in gantry<br />

and collimator 0 ◦ which is to check the fluence from the<br />

accelerator. The second irradiation is performed as if the<br />

phantom was the patient, and then each field is summed<br />

up to give a view of the complete dose to the<br />

patient/phantom. Then the measured and calculated<br />

dose distribution on the phantom is compared using γ<br />

evaluation and a visual comparison of the profiles. We<br />

use the criteria 3% and 3mm in our γ evaluation.<br />

In vivo QC<br />

For an in vivo system it is important to check initially<br />

that the system is stable and constant over time. This<br />

can be made that one irradiates the diodes 10-15 times<br />

and the spread should be within 0.5%. The<br />

measurements should be remade at different days for 1-<br />

2 weeks to see that the diodes are constant.<br />

3. Results<br />

We have an uptime of 97-98%. Using the in house made<br />

Excel sheet the physicist can follow the output from the<br />

daily QA check as can be seen in Fig. 6 for 4 MV<br />

photons and in Fig. 7 for 10 MV photons.<br />

Fig. 6. Per cent deviation from reference output in the daily<br />

morning QA for 4 MV photons.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!