05.08.2013 Views

Spot Vital Signs Service Manual - Frank's Hospital Workshop

Spot Vital Signs Service Manual - Frank's Hospital Workshop

Spot Vital Signs Service Manual - Frank's Hospital Workshop

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Service</strong> <strong>Manual</strong> Troubleshooting 31<br />

Table 8. Inaccurate Blood Pressure Readings<br />

Possible Cause Explanation and Corrective Action<br />

Incorrect reference Use the correct Korotkoff sound to determine diastolic blood pressure.<br />

Many listeners incorrectly equate diastolic blood pressure with the<br />

disappearance of sound only (phase 5). The Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> was<br />

developed using the American Heart Association recommendations, which state<br />

that phase 5 be used unless sound continues to 0 mmHg, in which case the<br />

change in the quality of sound (phase 4) is to be used.<br />

Deflate blood pressure cuff no faster than 3 mmHg per second.<br />

One of the major sources of error in auscultatory blood pressure measurement is<br />

deflating the blood pressure cuff too quickly. The American Heart Association<br />

recommends deflation no faster than 3 mmHg per second.<br />

Only use a sphygmomanometer that is calibrated.<br />

An uncalibrated sphygmomanometer may take inaccurate blood pressure<br />

measurements.<br />

Change in blood pressure<br />

between auscultatory reading<br />

and Welch Allyn <strong>Spot</strong> <strong>Vital</strong><br />

<strong>Signs</strong> reading<br />

Poor auscultatory sound<br />

recognition by observer<br />

Check blood pressure immediately prior to Welch Allyn <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> reading.<br />

Use higher quality stethoscope. Have a different observer check patient’s blood<br />

pressure.<br />

Note: Differences of up to 10 mmHg are considered normal and occur for a number of reasons including intra-patient<br />

blood pressure variability, observer hearing differences, and auscultatory deflation rate.<br />

Table 9. Cuff Inflation and Deflation with No Blood Pressure Reading Displayed (or Error Code in<br />

Display)<br />

Possible Cause Explanation and Corrective Action<br />

Leak in pneumatic system Ensure all blood pressure cuff attachments are tight. Carefully<br />

check for leaks in the blood pressure cuff, tubing, and pressure<br />

hose attached to the device.<br />

Arm movement during cycle Keep arm still during blood pressure cycle. Movement may cause<br />

inaccuracies from artifact.<br />

Blood pressure cuff tubing or pressure hose<br />

movement artifact<br />

Table 10. No Blood Pressure Cuff Inflation<br />

Do not contact blood pressure cuff tubing or pressure hose<br />

during blood pressure cycle. Movement may cause inaccuracies<br />

from artifact.<br />

Possible Cause Explanation and Corrective Action<br />

Connections between device and blood<br />

pressure cuff loose<br />

Table 11. Temperature Malfunction<br />

Check all connections (do not overtighten).<br />

Possible Cause Explanation Corrective Action<br />

Error code displayed Broken probe Replace probe. Consult <strong>Service</strong> <strong>Manual</strong>. Notify biomedical<br />

department or Welch Allyn Technical Support.<br />

Low temperature<br />

readings<br />

Improper probe<br />

placement<br />

Place probe in most posterior sublingual pocket when in<br />

Oral Mode. Verify patient has had nothing to eat or drink<br />

for 20 minutes.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!