<strong>Service</strong> <strong>Manual</strong> Troubleshooting 35<br />

Current test<br />

Noise levels<br />

Button test<br />

3. Connect to the 500 cc cylinder.<br />

4. Select Test > Calibration in the repair software. The dialog box displays the <strong>Spot</strong> <strong>Vital</strong><br />

signs manometer reading, battery reading, and valve and pump status.<br />

5. Push the Blood Pressure Start/Stop button to close the valve.<br />

6. Use the hand bulb to verify <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> is within calibration specification at the<br />

target pressures of 250±5 mmHg, 150 ±5 mmHg, 50 ±5 mmHg, and 0 mmHg with<br />

each reading within ±3 mmHg of the target pressure except for the 0 mmHg reading<br />

which is within ±1.0 mmHg.<br />

7. Press the Mode button until the LCD window displays bat.<br />

8. Set the power supply to 5.6 Vdc (+0.3/-0.0 Vdc). Verify the voltage reading meets the<br />

specification (see “Voltage calibration” on page 74).<br />

After completing this test, return the power suppy to 6.5 Vdc (+0/-0.25 Vdc).<br />

9. Select OK in the Test Calibration dialog box.<br />

1. Follow the steps in “Connections” on page 25.<br />

2. Power off <strong>Spot</strong> <strong>Vital</strong> signs and then power on in normal mode.<br />

3. Select Test > Current Levels in the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> Repair Software. Use the reading<br />

from the current meter and check the following current levels:<br />

a. Blank<br />

b. Back Light (Idle)<br />

c. Valve/Pump<br />

d. SpO 2 Mode<br />

The test result is a pass or fail result.<br />

4. Select Ok to exit the dialog box.<br />

1. Follow the steps in “Connections” on page 25.<br />

2. Select Test > Noise Levels in the <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> Repair Software and press the Test<br />

button to retrieve <strong>Spot</strong> <strong>Vital</strong> <strong>Signs</strong> internal pressure channel noise level. The test<br />

result is a pass or fail result (page 73).<br />

3. Select Ok to exit the dialog box.<br />

1. Follow the steps in “Connections” on page 25.