Sorting and Tolerance Checking<br />
12<br />
Sorting and Tolerance Checking<br />
In the sorting and tolerance checking mode, the display unit<br />
compares the displayed value with the programmed upper<br />
and lower sorting limits. The sorting and tolerance checking<br />
mode is enabled and disabled with operating parameter P17.<br />
Entering sorting limits<br />
Sorting limits are entered in operating parameters P18 and<br />
P19 (see “Operating Parameters”).<br />
Sorting signals<br />
The indicators and switching outputs at D-sub connection<br />
EXT (X41, see page 34) sort the display value into one of three<br />
classes.<br />
Display Meaning<br />
= Measured value is within sorting limits<br />
< Measured value is smaller than lower sorting limit<br />
> Measured value is greater than upper sorting limit<br />
Operating parameters for sorting and tolerance checking<br />
Sorting ON/OFF<br />
Lower sorting limit<br />
Part 1<br />
Part 2<br />
Part 3<br />
Upper sorting limit Example: Upper sorting limit = 26.02 mm<br />
Lower sorting limit = 26.00 mm<br />
gut Accept<br />
Nacharbeit Rework<br />
Ausschu§˜ Reject