RS-232-C/V.24 Interface (X31)<br />
34<br />
Measured value output after signal “STX” (Ctrl B)<br />
If the display unit receives the control character STX (Ctrl B)<br />
over the RS-232-C/V.24 interface (X31), it outputs the current<br />
measured value over the interface.<br />
➤ Transfer the control character Ctrl B over the RXD line of<br />
the RS-232-C/V.24 interface (X31).<br />
Characteristic times for measured value output<br />
Process Time<br />
Storage delay t1 ≤ 1ms<br />
Measured value output after t 2 ≤ 22 ms<br />
Regeneration time t 3 ≥ 0<br />
These times are prolonged if functions are active<br />
(for example, series of measurements with DIFF<br />
value display).<br />
Duration of measured value transfer<br />
t<br />
187 + (11 number of blank lines)<br />
D = [s]<br />
baud rate<br />
10 L%=17<br />
20 CLS<br />
30 PRINT "V.24/RS-232-C"<br />
40 OPEN "COM1:9600,E,7" AS#1<br />
50 PRINT #1, CHR$ (2);<br />
60 IF INKEY$""THEN 130<br />
70 C%=LOC(1)<br />
80 IF C%