Error Messages<br />

Display Problem<br />

Last measured value has not been output 1)<br />

The external device is not connected, no<br />

DSR signal (only displayed once!) 1)<br />

Data interface: Parity error or wrong transfer<br />

format 1)<br />

Incorrect input value<br />

Overflow due to external setting (value for<br />

P79 too high)<br />

Overflow trigger limit 1<br />

Overflow trigger limit 2<br />

Overflow lower sorting limit<br />

Overflow upper sorting limit<br />

The encoder signal is too weak. The scale<br />

may be contaminated. 1)<br />

The input frequency for this encoder input<br />

is too high. This can occur when the scale<br />

is moved too fast. 1)<br />

Internal counter overflow 1)<br />

Error during traverse of the reference<br />

marks1) 1) These errors are important for the attached device. The<br />

error signal (pin 19) at D-sub connection EXT is active.<br />

Display Problem<br />

If these errors persist, contact your<br />

service agency.<br />

Offset compensation values for encoder<br />

signals have been erased: contact your<br />

service agency.<br />

Check the operating parameters. If this<br />

error persists, contact your service<br />

agency.<br />

Other error displays<br />

If all decimal points light up, the measured value is too<br />

great or too small:<br />

➤Set a new datum<br />

or<br />

➤ Traverse back.<br />

If all sorting signals light up, the upper sorting limit is<br />

smaller than the lower limit:<br />

➤ Change operating parameters P18 and/or P19.<br />

To clear error messages:<br />

When you have removed the cause of the error:<br />

➤ Clear the error message with the CL key.<br />

Delete error messages ERROR 80, 83, 84, 86.<br />

Switch device off!<br />

Error Messages<br />

15