RS-232-C/V.24 Interface (X31)<br />
32<br />
Operating parameters for measured value output<br />
Parameter Function<br />
Baud rate<br />
Number of additional blank lines for<br />
measured value output<br />
Display freeze during measured value output<br />
In operating parameter P23, you can specify how the<br />
measured value output signal will affect the display unit.<br />
Display freeze during measured value outputP23<br />
Concurrent display, no display freeze:<br />
The display value is the current measured value<br />
Frozen display: Display is stopped<br />
(frozen) and updated by every<br />
measured value output signal<br />
Frozen/concurrent display: Display is<br />
frozen as long as a measured value output<br />
signal is present<br />
To output measured values with the PRINT function:<br />
➤ Press MOD repeatedly,<br />
until the indicator PRINT blinks.<br />
➤ Start measured value output with ENT.<br />
Duration of measured value transfer<br />
187 + (11 number of blank lines)<br />
tD = [s]<br />
baud rate<br />
Indicator preselection<br />
Operating parameter P86 allows you to define which indicator<br />
is displayed first when MOD is pressed.