The Setup Screen displays the factory default settings the<br />

first time the dosimeter is used. Thereafter, stored settings<br />

are displayed.<br />

Measurement Mode<br />

NOTE: Exchange rate, criterion level and threshold level<br />

determine how noise dose is defined and computed, and<br />

will impact the dose results. The default settings correspond<br />

to NIOSH (1998) recommendations, which are more<br />

conservative (more protective) than the OSHA criteria.<br />

NIOSH (default)<br />

Exchange rate: 3 dB<br />

Criterion level: 85 dB<br />

Threshold level: 75 dB<br />

OSHA<br />

Exchange rate: 5 dB<br />

Criterion level: 90 dB<br />

Threshold level: 80 dB<br />

Custom<br />

Exchange rate: 3, 4 or 5 dB<br />

Criterion level: 70 to 95 dB, in 1-dB increments<br />

Threshold level: 65 to 90 dB, in 1-dB increments<br />

41